Functions |
Range |
Resolution |
Accuracy |
BT508 |
BT510 |
BT520 |
BT5213 |
Battery Resistance / Strap Resistance1 |
3 mΩ |
0.001 mΩ |
1% + 8 |
● |
● |
● |
● |
30 mΩ |
0.01 mΩ |
0.8% + 6 |
● |
● |
● |
● |
300 mΩ |
0.1 mΩ |
0.8% + 6 |
● |
● |
● |
● |
3000 mΩ |
1 mΩ |
0.8% + 6 |
● |
● |
● |
● |
Vdc |
6 V |
0.001 v |
0.09% + 5 |
● |
● |
● |
● |
60 V |
0.01 v |
0.09% + 5 |
● |
● |
● |
● |
600 V |
0.1 v |
0.09% + 5 |
● |
● |
● |
● |
1000 v |
1V |
0.09% + 5 |
|
|
|
● |
Vac (45 Hz to 500 Hz with 800 Hz filter) |
600 v |
0.1 v |
2% + 10 |
|
● |
● |
● |
Frequency (displayed with Vac and Aac)2 |
500 Hz |
0.1 Hz |
0.5% + 8 |
|
● |
● |
● |
AC voltage ripple (20KHz Max) |
600 mV |
0.1 mV |
3% + 20 |
|
● |
● |
● |
6000 mV |
1 mV |
3% + 10 |
|
● |
● |
● |
Adc/Aac (with accessory Fluke i410) |
400 A |
1 A |
3.5% + 2 |
|
|
|
● |
Temperature |
0°C to 60°C |
1°C |
2°C (4 °F) |
|
|
|
● |
Meter mode |
999 records for each measurement position with time stamp |
Sequence mode |
Up to 100 profiles and 100 profile templates (Each profile stores up to 450 batteries) with time stamp |
1 The measurement is based on ac injection method. The injected source signal is < 100 mA, 1 kHz.
2 Trigger level VAC: 10 mV, Aac: 10 A
3 Not available in all regions |
Measurement Modes |
BT508 |
BT510 |
BT520 |
BT521 |
Resistance (mΩ) |
● |
● |
● |
● |
Battery voltage |
● |
● |
● |
● |
Voltage dc |
● |
● |
● |
● |
Voltage ac and frequency (Hz) |
|
● |
● |
● |
Ripple volt |
|
● |
● |
● |
Temperature of negative battery post |
|
|
|
● |
DC and ac current (and frequency) |
|
|
|
● |
DMM mode |
|
● |
● |
● |
Sequence mode |
|
● |
● |
● |
Discharge measurement mode |
|
● |
● |
● |
Automatic measurement save |
|
● |
● |
● |
Wireless communication |
|
|
|
● |
Memory view |
|
● |
● |
● |
General Specifications |
Size (HxWxD) |
22 cm x 10.3 cm x 5.8 cm (9 in x 4 in x 2 in) |
Weight |
850 g (1.9 lb) |
Screen dimensions |
7.7 cm x 5.6 cm (3 in x 2.2 in) |
Interface |
USB mini |
Warranty |
3 years |
Environment Specifications |
Operating temperature |
0°C to 40°C |
Storage temperature |
-20°C to 50°C |
Lithium-ion battery charging temperature |
0°C to 40°C |
Operating humidity |
Non-condensing (10°C) |
<=80% RH (at 10°C to 30°C) |
<=75% RH (at 30°C to 40°C) |
Operating altitude |
Sea level to 2000 meters |
Storage altitude |
Sea level to 12,000 meters |
IP rating |
IP40 |
Radio |
FCC Class A |
Vibration requirements |
MIL-PRF-28800F: Class 2 |
Drop test requirements |
1 meter |
Temperature coefficients |
Add 0.1 x specified accuracy for each degree C above 28°C or below 18°C |
Safety compliance |
600 V CAT III |
EMC |
IEC 61326 |
ROHS |
China, Europe |
Protection class 2 |
Pollution Degree II |
Battery compliance |
UN38.3 |
UL2054 |
IEC62133 |
2G per IEC68-2-26, 25G, and 29 |
|
|
|
|