- ATEX II 1G Ex ia IIC T4 Compliant.
- Built-in pressure / vacuum hand pump, with fine adjust vernier and bleed valve.
- 30, 100, and 300 psi ranges (2, 7, and 20 bar)
- Pressure measurement to 0.05% of full span, using an internal pressure sensor.
- Pressure measurement to 200 bar using any of eight intrinsically safe Fluke 700PEx Pressure Modules.
- Wide range of selectable measurement units for pressure.
- Current measurement with 0.02% accuracy and 0.001 mA resolution.
- Pressure switch test function.
- Min-Max Hold functions.
- Compact size and weight.
- Simple, push-button user interface.
- Rugged and reliable, ideal for field use.
Measurement Accuracy |
Pressure |
Fluke 718Ex 30G |
|
Range: |
-830 mbar to 2 bar (internal sensor) |
|
Over pressure: 3x FS |
Resolution: |
0.0001 bar |
Accuracy: |
0.05% FS |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
Fluke 718Ex 100G |
|
Range: |
-830 mbar to 7 bar (internal sensor) |
|
Over pressure: 2x FS |
Resolution: |
0.001 bar |
Accuracy: |
0.05% FS |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
Fluke 718Ex 300G |
|
Range: |
-830 mbar to 20 bar (internal sensor) |
|
Over pressure: 375 psi, 25 bar |
Resolution: |
0.0001 bar |
Accuracy: |
0.05% FS |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: for use with non-corrosive gasses |
Fluke 718Ex series |
|
|
via pressure module connector |
Range: |
29 pressure modules, 2.5 mbar to 700 bar |
|
Over pressure per pressure module specs |
Resolution: |
Per pressure module specs |
Accuracy: |
To 0.025% of full span, per pressure module specs |
Functions: |
Zero, Min, Max, Hold, Damp |
|
Note: media compatibility per pressure module specs |
Source Pressure, built in pump |
|
Range: |
-820 mbar to full scale |
Note: |
Supported Pressure Units PSI, in. H2O (4 °C), in. H2O (20 °C), kPa, cm H2O (4 °C), |
|
cm H2O (20 °C), bar, mbar, kg/cm², mmHg, in Hg |
|
Technical Data |
Measure Accuracy: |
Range: |
0-24 mA |
Resolution: |
0.001 mA |
Accuracy: |
0.02% rdg ± 2 counts |
|
Display |
|
LCD, 5 digit pressure and current simultaneous |
|
Environmental Specifications |
Operating Temperature |
|
Storage Temperature |
|
Humidity (Without Condensation) |
|
95% (10ºC to 30ºC) |
|
75% (30ºC to 40ºC) |
|
45% (40ºC to 50ºC) |
|
35% (50ºC to 55ºC) |
|
Operating Altitude |
|
Safety Specifications |
Safety |
|
ATEX II 1G Ex ia IIC T4 compliant |
|
CSA Certification I.S. Class I, |
|
Div 1 Groups A-D , T4 |
|
Mechanical & General Specifications |
Size |
|
216 mm L x 94 mm W x 66 mm H (with holster) |
|
Weight |
|
Approx. 992 g (with holster) |
|
Warranty |
|
Three years, one year for pump |
|
Shock |
|
Vibration |
|
Battery type |
|
One 9-volt alkaline battery ANSI/NEDA 1604A or IEC 6LR619V |
|
Battery life: 4 to 20 hours, depending on functions used |
|
|