OVERVIEW & SPECIFICATION
- ouch Panel with Color LCD Display
-
9 Different Measurement Network to Simulate the Resistance of Human Body
-
8 Different Types of Leakage current for Medical & General Electrical Equipment
-
Various Leakage Current Measuring Mode:DC/AC/AC+DC/AC Peak
-
Upper & Lower Limitation for PASS/FAIL Judgment
-
Meter Mode with SELV/CONV Function
-
50 Preset Test Conditions Conform to IEC 60990; 30 Sets Memories for Customer Defined
-
Various Standard Interfaces:RS232/GPIB/USB Host & Device/EXT I/O
General | Display | 320 x 240 Color TFT LCD |
Operation | Touch panel | |
Measurement Device | 9 kinds | |
DC | Ranges | 50uA/500uA/5mA/25mA |
Resolution | 0.01uA/0.1uA/1uA/10uA | |
Accuracy | ±1.0% fs/±(0.2% rdg+3dgt) | |
AC/AC+DC | Ranges | 50uA/500uA/5mA/25mA |
Resolution | 0.01uA/0.1uA/1uA/10uA | |
Accuracy | ±2.0% fs/±(2.0% rdg+6dgt) — 10Hz<f≦100kHz | |
±2.0% fs/±(2.0% rdg+10dgt) — 100kHz<f≦1MHz | ||
AC peak | Ranges | 500uA/1mA/10mA/75mA |
Resolution | 0.1uA/1uA/10uA/100uA | |
Accuracy | ±4.0%fs/±2.5%fs/±(2.0% rdg+6dgt) — 20Hz<f≦1kHz |
|
±5.0%fs/±5.0%fs/±(5.0% rdg+10dgt) — 1kHz<f≦10kHz |
||
Meter Mode | AC/DC/AC+DC | 10.0~300.0V |
AC peak | 15.0~430.0V | |
SELV function | Available | |
EUT (V/I Check) | Voltage range | 85V ~ 300V |
Current range | 0.5A ~ 10A | |
Memory | Empty sets | 30 sets |
Preset sets | 50 sets (conform to IEC 60990) | |
Interfaces | Front panel | USB host |
Rear panel | RS-232C/GPIB/EXT I/O/USB device | |
Power Source | For GLC-9000 | AC 100V/120V/220V/230V±10%,50/60Hz |
For EUT | AC 85V~250V,50/60Hz | |
Dimensions | 330 (W) x 150 (H) x 350 (D) mm | |
Weight | Approx. 5kg |
GLC-9000
OVERVIEW & SPECIFICATION
- ouch Panel with Color LCD Display
-
9 Different Measurement Network to Simulate the Resistance of Human Body
-
8 Different Types of Leakage current for Medical & General Electrical Equipment
-
Various Leakage Current Measuring Mode:DC/AC/AC+DC/AC Peak
-
Upper & Lower Limitation for PASS/FAIL Judgment
-
Meter Mode with SELV/CONV Function
-
50 Preset Test Conditions Conform to IEC 60990; 30 Sets Memories for Customer Defined
-
Various Standard Interfaces:RS232/GPIB/USB Host & Device/EXT I/O
General | Display | 320 x 240 Color TFT LCD |
Operation | Touch panel | |
Measurement Device | 9 kinds | |
DC | Ranges | 50uA/500uA/5mA/25mA |
Resolution | 0.01uA/0.1uA/1uA/10uA | |
Accuracy | ±1.0% fs/±(0.2% rdg+3dgt) | |
AC/AC+DC | Ranges | 50uA/500uA/5mA/25mA |
Resolution | 0.01uA/0.1uA/1uA/10uA | |
Accuracy | ±2.0% fs/±(2.0% rdg+6dgt) — 10Hz<f≦100kHz | |
±2.0% fs/±(2.0% rdg+10dgt) — 100kHz<f≦1MHz | ||
AC peak | Ranges | 500uA/1mA/10mA/75mA |
Resolution | 0.1uA/1uA/10uA/100uA | |
Accuracy | ±4.0%fs/±2.5%fs/±(2.0% rdg+6dgt) — 20Hz<f≦1kHz |
|
±5.0%fs/±5.0%fs/±(5.0% rdg+10dgt) — 1kHz<f≦10kHz |
||
Meter Mode | AC/DC/AC+DC | 10.0~300.0V |
AC peak | 15.0~430.0V | |
SELV function | Available | |
EUT (V/I Check) | Voltage range | 85V ~ 300V |
Current range | 0.5A ~ 10A | |
Memory | Empty sets | 30 sets |
Preset sets | 50 sets (conform to IEC 60990) | |
Interfaces | Front panel | USB host |
Rear panel | RS-232C/GPIB/EXT I/O/USB device | |
Power Source | For GLC-9000 | AC 100V/120V/220V/230V±10%,50/60Hz |
For EUT | AC 85V~250V,50/60Hz | |
Dimensions | 330 (W) x 150 (H) x 350 (D) mm | |
Weight | Approx. 5kg |
Leakage Current TestersSafety Testers
- AC + DC Power Sources
- AC Power Source
- AC Power Sources
- Benchtop LCR Meters
- DC Electronic Loads
- DC Power Supplies
- Handheld LCR Meters
- Handheld Oscilloscopes
- LCR Meters
- Leakage Current Testers
- Multiplex Scanner Box
- Non-Programmable & Multiple Channel
- Non-Programmable & Single Channel DC Power Supplies
- Oscilloscopes
- Programmable & Single Channel DC Power Supplies
- Real-Time & Digital Storage
- Safety Testers
- Safety Testers (Hi-Pot Tester)